BB-100 Standard
MDTHYQ1_0211
feature & parameter
Main technical parameter
1 Power Supply: ~ 220V, 50/60Hz
2 Power input: 650VA(including power output socket :200VA)
3 Operation environment conditions:
- An ambient Temperature: 20℃- 30℃
- An ambient Humidity: 30% RH - 75% RH
- An ambient air pressure: 700 hPa -1060 hPa
- An ambient air velocity less the 0.3m/s
4 Control temperature range:
- Air temperature : 25℃-37℃, 37.1℃-38℃
5 The warm-up time: 45 minutes
6 The function of alarm:
6.1 The over temp of air temp : 38℃ (cut off heating power, together with audible and visual alarm)
6.2 The second thermal cut-out temperature: 40℃ (cut off heating power, together with audible and visual alarm)
6.3 The alarm for deviation:
- Air temp: ±3.0℃ (Heater is disconnected simultaneously at +3℃) audible and visual alarm)
6.4 Alarm for Sensors:
When the skin (or air) temperature sensor is shut off or short circuit,cut off the heater,give out alarm with sound and light .
6.5Alarm for power failure
The power fails,give out alarm with sound and light.
6.6 Alarm for fan failure
The fan stops working or has fault, give out alarm with sound and light and cut off the heater power.
6.7 Max CO2-concentration is 0.4%
6.8 Normal working, the whole equipment noise in the baby cabin less then 55dB(A)
6.9 Transport and storage
Environment temperature: -10℃to +55℃
Relative humidity ≤95%RH
Atmospheric pressure range: 500hPa-1060hpa
feature & parameter
Main technical parameter
1 Power Supply: ~ 220V, 50/60Hz
2 Power input: 650VA(including power output socket :200VA)
3 Operation environment conditions:
- An ambient Temperature: 20℃- 30℃
- An ambient Humidity: 30% RH - 75% RH
- An ambient air pressure: 700 hPa -1060 hPa
- An ambient air velocity less the 0.3m/s
4 Control temperature range:
- Air temperature : 25℃-37℃, 37.1℃-38℃
5 The warm-up time: 45 minutes
6 The function of alarm:
6.1 The over temp of air temp : 38℃ (cut off heating power, together with audible and visual alarm)
6.2 The second thermal cut-out temperature: 40℃ (cut off heating power, together with audible and visual alarm)
6.3 The alarm for deviation:
- Air temp: ±3.0℃ (Heater is disconnected simultaneously at +3℃) audible and visual alarm)
6.4 Alarm for Sensors:
When the skin (or air) temperature sensor is shut off or short circuit,cut off the heater,give out alarm with sound and light .
6.5Alarm for power failure
The power fails,give out alarm with sound and light.
6.6 Alarm for fan failure
The fan stops working or has fault, give out alarm with sound and light and cut off the heater power.
6.7 Max CO2-concentration is 0.4%
6.8 Normal working, the whole equipment noise in the baby cabin less then 55dB(A)
6.9 Transport and storage
Environment temperature: -10℃to +55℃
Relative humidity ≤95%RH
Atmospheric pressure range: 500hPa-1060hpa