MDTHYQ1_03608
MDTSZRF75527460837
MDTHYQ1_03608
1) Temperature range: 32.0°C~42.9°C [89.6F~109.2F]
2) Accuracy:
35.0°C or less ±0.2°C [95.0F or less, ±0.4F]
35.1℃~39.0℃ ±0.1℃ 【95.2F~102.2F,±0.2F】
39.1°C or more, ±0.2°C [102.4F or more, ±0.4F]
3) Resolution: 0.1°C/0.1F
4) Auto Shutdown Time: Auto OFF within 10 minutes after °C/F stops flashing.
5) Center Point Display: Fixed display 37.0°C (98.6°F) for 1 second.
6) If the measurement temperature is below 32.0 °C (89.6 F), it will show L0 °C (L0F), and above 42.9 °C (109.2 °F)
then H°C (H°F) is displayed.
7) Temperature measurement stability time: if the temperature does not change again within 16 seconds, the °C/°F will stop flashing, and there will be an ALARM "B" sound, indicating that the temperature is stable
If the temperature rises again, the rising value is displayed, but the °C/°F will not flash again, and the ALARM will not sound (i.e. the ALARM will only sound the first stable temperature).
8) Current consumption: OFF< 0.5μA, ON< 150μA (Power supply: DC1.5V)
9) "B" ALARM: When you press the ON/OFF button, there will be a short sound of Bi, the temperature value will not change within 16 seconds, and the temperature will be measured
After stabilization, 4 Bi~Bi sounds appeared.
10) "M" memory function: press ON/OFF SW, the last measured temperature will be displayed directly, and an M symbol will appear on the right side of the LCD.
11) Low voltage function: when the battery voltage is too low, the symbol of "□" appears, France and PTB: 1.23V±0.05V (RL=470KΩ) General low voltage: 1.10V~1.25V
1) Temperature range: 32.0°C~42.9°C [89.6F~109.2F]
2) Accuracy:
35.0°C or less ±0.2°C [95.0F or less, ±0.4F]
35.1℃~39.0℃ ±0.1℃ 【95.2F~102.2F,±0.2F】
39.1°C or more, ±0.2°C [102.4F or more, ±0.4F]
3) Resolution: 0.1°C/0.1F
4) Auto Shutdown Time: Auto OFF within 10 minutes after °C/F stops flashing.
5) Center Point Display: Fixed display 37.0°C (98.6°F) for 1 second.
6) If the measurement temperature is below 32.0 °C (89.6 F), it will show L0 °C (L0F), and above 42.9 °C (109.2 °F)
then H°C (H°F) is displayed.
7) Temperature measurement stability time: if the temperature does not change again within 16 seconds, the °C/°F will stop flashing, and there will be an ALARM "B" sound, indicating that the temperature is stable
If the temperature rises again, the rising value is displayed, but the °C/°F will not flash again, and the ALARM will not sound (i.e. the ALARM will only sound the first stable temperature).
8) Current consumption: OFF< 0.5μA, ON< 150μA (Power supply: DC1.5V)
9) "B" ALARM: When you press the ON/OFF button, there will be a short sound of Bi, the temperature value will not change within 16 seconds, and the temperature will be measured
After stabilization, 4 Bi~Bi sounds appeared.
10) "M" memory function: press ON/OFF SW, the last measured temperature will be displayed directly, and an M symbol will appear on the right side of the LCD.
11) Low voltage function: when the battery voltage is too low, the symbol of "□" appears, France and PTB: 1.23V±0.05V (RL=470KΩ) General low voltage: 1.10V~1.25V