Name | Disposable Infusion Pump |
Type | CBI/CBI+PCA 60/100/150/200/275/300ml |
Infusion volume | not less than 85% of nominal capacity |
infusion pump structure | Protection bottle /reservoir /skull protector /Infusion tube /Major catheter for transfusion/ Stanching clip /PCA fiter / PCA button /Ruhr joint |
Packing for 60mL,100mL 28*18cm | 1 piece/Bag 10 pieces/box 100 pieces/carton Box size:34*21*15cm Carton size: 78*35.5*44cm |
Packing for 150ml/200ml /275ml/300ml 33*18cm | 1 piece/Bag 5 pieces/box 50 pieces/carton Box size:34*21*15cm Carton size: 78*35.5*44cm |
Leading Time | 31 days after payment |
Features
High Accuracy & Safety*Stable Flow Rate Promotes Patient Mobility Fast Filling
Ambulatory Design Doesn't require continuous patient monitoring
1.2 pm high precision air & particle fitter.
Non Breakable" Body Available in Multiple
specifications.including single flow rate, 4-flow rate and 15 flow rate pumps (with or without PCA) suitable to meet all clinical
requirements One way injection port Comes with Special Medicine Return Device.Used for patients requiring intravenous, percutaneous, subcutaneous, intraoperative sites or epidural administration of medications; Medications include Analgesics, Chemotherapeuties, Antibiotics, Antiviral drugs and Infusion for re-hydration; There are various volume and flow rate to match each other and meet the different clinical requirements; PCA device can make the patient to add the drug by herself himself according to pain situation; There are Multi rate and UV type, including 15 flow rate, portable and safety;
Silicone reservoir allows air to dissipate automatically; There are different drug filters, including 5μ, 1.2μ and 0.2μ;
Accuracy of flow rate can be controlled +10%; DEHP fee and Latex free.
Nominal Capacity | Continually dosing volume | Self-control dosing volume | Interval |
60ml | 1.0ml/hour 1.5ml/hour 2.0ml/hour 2.5ml/hour 3.0 ml/hour 4.0ml/hour 5.0ml/hour 6.0ml/hour | 0.5ml/1ml/2ml/time | 10min/15min |
100ml | 0.5ml/1ml/2ml/time | 10min/15min |
150ml | 0.5ml/1ml/2ml/time | 10min/15min |
200ml | 0.5ml/1ml/2ml/time | 10min/15min |
275ml | 0.5ml/lml/2ml/time | 10min/15min |
300ml | 0.5ml/lml/2ml/time | 10min/15min |
Notices
1. Use within the period of validity of sterilization. and package in damaged condition is forbidden for use.
2. Fluid injected into reservoir must be diluted by rational propotion, it is forbidden to inject undiluted fluid directly in vivo of the sufferer.
3. Inject an specified dosage into pump, and no excess is allowable for fear of affecting the use effect or resulting in rupture of sacculus.
4.The marked flow rate is tested at room temperature 23℃ ±2℃ with the medium of distilled water, The flow rate may vary due to the difference of medicine recipe, concentration, temperature, hanging height of main body.viscosity of fluid and injection pressure, which shall be observed with real time such that grasping the individual differences of patients and understand the demands of each sufferer.
5. In 1 to 2 hours of inception of product application, the flow rate may be faster than the marked (but still falling in the scope as set in product standards), which is determined by physical performances of silica capsule.if any gases found in silica capsule, they could be exhausted after several decades of minutes under the action of silica gel.
6.This product has been subject to oxirane sterilization, with validity of 5 years.
7.This product is only for Disposable, it shall be melt down immediately after application.
8.Your must read and understand the use method and functions of the product to ensure a safely and effeetive use.
9.The fluid shall be prepared by combining the flow rate set for this product and the needs of sufferer to prevent the sufferer from unnecessary pain or medical accident due to improper concentration or mixture ratio.